Test and application of compact digital X-ray imaging system
-
摘要:研制了一套采用X射线敏感CCD作为成像探测器的微型数字成像系统并用于牙科诊断。应用牙齿模型与分辨率测试卡对系统的辐射剂量和分辨率进行了测试,采用分段线性校正对CCD与转换屏耦合下的响应不均匀性进行校正,并通过计算线性相关系数辨别瑕疵像元以避免误判。测试了系统在高分辨率、小工作幅面X射线检测方面的应用效果。结果显示,系统分辨率高于10 lp/mm,在获得相近对比度图像的条件下,所需的辐射剂量仅为胶片的10%,图像非均匀性相对校正前可降低27.5%。得到的结果表明,系统满足高分辨率、低辐射剂量的使用要求,在微小物体X射线检测方面具有较大的应用潜力。Abstract:A compact digital X-ray imaging system which adopts an X-ray sensitive CCD as an imaging detector was introduced for the dental imaging. A tooth model and a resolution testing card were used to test the revolution and radiographic dosage, respectively, and the subsection linearity correction was used to calibrate the nonuniformity of coupling of CCD and switch screen. In order to identify the bad pixel, a linear correlation coefficient was calculated to avoid the misjudgement. The imaging system was applied to an X-ray detection required a high resolution and a small work-area and the results show that the resolution of imaging system is higher than 10 lp/mm, radiographic dosage is less than 10% in contrast with that of X-ray film, and the calibrated nonuniformity of image has reduced by 27.5%. Obtained results prove that the system can satisfy the system requirements of high-revolution, low- radiographic dosage, and has a strong applied potential in X-ray detection for small objects.
-
[1] ROWLANDS J A. The physics of computed radiography[J].Institute Physics Publishing,Phys. Med. Biol.,2002,47(2):123-166. [2] HAKAN G. Image quality VS radiation dose for a flat-panel amorphous silicon detector:a phantom study[J].Eur Radiol,2001,11:1704-1709. [3] FRANK F,JENS R. Flat panel digital radiography compared to storage phosphor computed radiography:assessment of dose versus image quality in phantom studies[J].Invest Radiol,2002,37:609-614. [4] GRANFORS P R. Performance characteristics of an amorphous silicon flat panel X-ray imaging detector[J].SPIE,1999,3659:480-490. [5] LEE D L. Improved imaging performance of a 14 X 17-inch direct radiography system using Se/TFT detector[J].SPIE,1998,3336:14-23. [6] 郑玉权,王慧. 微型X射线数字成像系统[J]. 光学 精密工程 ,2008,16(4):591-597. ZHENG Y Q,WANG H. A compact digital X-ray imaging system[J].Opt. Precision Eng.,2008,16(4):591-597.(in Chinese) [7] 李云飞,司国良,郭永飞. 科学级CCD相机的噪声分析及处理技术[J]. 光学 精密工程 ,2005,13(Supp.):158-163. LI Y F,SI G L,GUO Y F. Noise analyzing and processing for scientific grade CCD[J].Opt. Precision Eng.,2005,13(Supp.):158-163.(in Chinese) [8] 白延柱,金伟其.光电成像原理与技术[M]. 北京:北京理工大学出版社,2006. BAI Y Z,JIN W Q.Principle and Technology of Photo Electricity Imaging[M]. Beijing:Beijing Technology University Press,2006.(in Chinese) [9] 马跃洲,艾维平,张昌青. 平板探测器X射线数字成像质量[J]. 兰州理工大学学报 ,2007,33(2):20-24. MA Y ZH,AI W P,ZHANG CH Q. X-ray digital imaging quality of flat panel detector[J].J. Lanzhou University Technol.,2007,33(2):20-24.(in Chinese) [10] 刘光达,赵立荣. 基于最小均方误差原理的医学X光影像滤波阈值选择[J] 光学 精密工程 ,2001,9(1):47-50. LIU G D,ZHAO L R. Threshold selection method for medical X-ray images filter based on minimum even-square error[J].Opt. Precision Eng.,2001,9(1):47-50.(in Chinese) [11] 邓超,张涛,姚清华. 应用小波神经网络处理CCD图像噪声[J]. 光学 精密工程 ,2008,16(2):345-351. DENG C,ZHANG T,YAO Q H. Application of wavelet neural network in removing CCD noise of digital images[J].Opt. Precision Eng.,2008,16(2):345-351.(in Chinese) [12] 程万胜,赵杰,蔡鹤皋. CCD像素响应非均匀性的校正方法[J]. 光学 精密工程 ,2008,16(2):314-318. CHENG W SH,ZHAO J,CAI H G. Correction method for pixel response nonuniformity of CCD[J].Opt. Precision Eng.,2008,16(2):314-318.(in Chinese)
点击查看大图
计量
- 文章访问数:4627
- HTML全文浏览量:458
- PDF下载量:2095
- 被引次数:0